GE四排螺旋CT机架角度故障
GE多排螺旋CT机架角度故障 机型:GE LightSpeed Ultra 故障:打角度时,机架运行缓慢,间断停止 故障信息:File: emcTouchSensor.c Line: 125 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 210/3/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:18:25:260 Log Series: 167 The Gantry FORWARD Touch Sensor Error Console Pushbutton Remote Tilt is DISABLED Use Gantry Controls to Tilt the Gantry
Current Position 12498 (.01 mm). Elev Dir: 0 (0 = Up 1 = Down) (Elev Cmd: 1 TouchSensor: 0x4000). 故障分析: 根据故障信息可以大致分析是Gantry Tilt Touch Sensors故障,打开机架左右侧盖,仔细检查四个GANTRY OPERATION BOARD,信号线及电源线连接头紧固,依次拆下各板检测,发现后板Gantry Tilt Touch Sensors连接不良,处理后正常
|